KTL1600 atmosphere tube furnace

Product name: KTL1600 atmosphere tube furnace

Model number: KTL1600

Minimum starting quantity: 1

Product Summary:
KTL1600 Air tube furnace with 1800 type silicon molybdenum rod for heating elements, using double shell structure and yu electric 40 program temperature control system, phase trigger, silicon control, furnace with 1800 type alumina crystal fiber material, double furnace shell with air cooling system, can quickly lift temperature, using 99 jade tube, both ends with stainless steel flange seal, control the intake flow with a float flowmeter.

product description
KTL1600 Degree introduction furnace tube furnace:

KTL1600 Degree of tubular furnace the tubular furnace with silicon molybdenum rod for heating elements, the double shell structure and temperature control instrument, can 30 section program temperature control, phase trigger, control, silicon furnace polycrystalline fiber material furnace, furnace tube both ends can be stainless steel flange sealing, stainless steel flange installed with air nozzle, valve and pressure gauge, vacuum vacuum can reach 10-3 Pa, the furnace has temperature equilibrium, low surface temperature, lifting temperature rate is fast, energy saving and other advantages, It is an ideal product for universities, scientific research institutes, industrial and mining enterprises to do high temperature atmosphere sintering, atmosphere reduction, CVD experiment and vacuum annealing.

The NO.1 heating element
(1) The thermal end thickness of silicon molybdenum rod D1 is 7mm, 6MM used by general manufacturers, the heating end is thick, not easy to bend, the surface load of the rod is low, and long service life.
(2) The use of hot bending technology, uniform resistance and long service life
(3) The length of Lu cold end is 210mm, the 400X300X300 furnace is 260, generally manufacturers use 180mm, the cold end long furnace roof temperature is low.
(4) The surface oxidized amorphous film is dense, which can better protect the silicon molybdenum rod.

NO.2 air switch and electric leakage protection
When the circuit is overcurrent or leakage, the empty meeting is automatically disconnected.
NO.3 Connect to a computer (buy)
KTL1600 Atmosphere tube furnace can be equipped with communication interface and software, can directly through the computer control the parameters of the furnace, and can observe from the computer furnace PV and SV temperature value and instrument operation, the actual heating curve computer real-time drawing, and can save the temperature data of each moment, can be out at any time.
NO.4 furnace
(1) vacuum suction filter formed alumina polycrystalline fiber curing furnace.
(2) Using Japanese technology to shape.
(3) the furnace material has low heat capacity, low heat conduction coefficient, so that less heat dissipation, good heat preservation, high reflectivity. Temperature field equilibrium.
(4) furnace alumina wire thin and long, high temperature resistance, heat rise and shrinkage ability.
(5) The furnace surface is smooth, not easy to drop powder.

NO.5 optional UL certified panel (all imported appliances with UL certification)
The NO.6 triggers
(1) Phase shift trigger (with soft start and flow limiting function)
(2) Current limiting function: no matter how to adjust the heating time, the secondary current of our electric furnace after the transformer is limited to 170A, with limited protection of the equipment. Non-current limiting trigger, if the heating rate is set correctly or the current is too high at low temperature, the equipment will trip or burn out.
(3) Soft start function: the current will slowly increase when the startup, and the impact on the heating element is small, which is conducive to the extension of the life of the heating element
(4) Protection function: if the current exceeds the limit current, the trigger will automatically cut off the trigger function to cut off the current protection electric furnace.
NO.7 UI certified transformer
(1) Transformer cable is pure copper
(2) The transformer sound is small
(3) Transformer temperature and body is low
NO.8 A variety of flanges are available
The NO.9 certification

Control mode: adopt temperature control instrument (standard)
1.Intelligent PID adjustment of temperature control in section 30.
2.with over temperature protection when the electric furnace heating circuit automatically over temperature or break off, (when the electric furnace temperature exceeds 1200 degrees or the thermocouple burned off, the AC relay on the main circuit will automatically disconnected, the main circuit disconnected, the panel ON lamp off, OFF light, limited protection electric furnace).
3.With the function of power off protection, that is, when the power is cut off and the power is started, the program does not rise from the starting temperature, but from the furnace temperature when the power is cut off.
4.The instrument has the function of temperature self-setting

Control accuracy: + / -1℃
Trigger: Phase shift trigger (with soft start and flow limit)
Current limiting function: no matter how to adjust the heating time, the secondary current of our electric furnace after the transformer is limited to 160A. With limited protection of the equipment, no matter how you operate, the equipment will not be burned out.
Soft start function: the current will slowly increase when starting up, and the impact on the heating element is small, which is conducive to the extension of the life of the heating element
Electrical appliances: Zhejiang Chint
SiriCR: 106 / 16E Simenon, Germany
Maximum temperature: 1,600℃
Rated temperature: 1,550℃
Heating rate: 20℃ / Min (can be modified upon request)
Recommended heating rate: 10℃ / Min
Thermocouple model: Type B
Heating section length: 300mm