KBF1700-HAF hydrogen atmosphere furnace

Product name: KBF1700-HAF hydrogen atmosphere furnace
Model number: KBF1700-HAF

Product Summary:
KBF1700-HAF hydrogen atmosphere furnace with 1800 molybdenum wire as the heating element, using double shell structure and Japan island power 40 section program temperature control system, phase trigger, silicon control, furnace using aluminum oxide polycrystal fiber material, double furnace shell equipped with air cooling system, can quickly lift temperature, using the whole sealing, cover plate sealing using silicone mud, silicone pad and water cooling system, gas after the flow meter, multiple washing furnace inlet, outlet with combustion mouth, can pre-vacuum and hydrogen gas, argon gas, nitrogen, oxygen, carbon monoxide, ammonia decomposition gas

product description
KBF1700-HAF hydrogen atmosphere furnace is described in detail:

KBF1700-HAF hydrogen atmosphere furnace with type 1800 molybdenum wire as the heating element, Using the double-layer shell structure and the Yudian 30-section program temperature control system, Phase shift triggered, SCR controlled, The furnace is made of alumina polycrystal fiber material, Double furnace housing with air cooling system, Can quickly lift and lower the temperature, Using shell integral sealing, cover sealing with silicone mud, furnace door with silicone pad and water cooling system, After the flow meter, There are many washing furnace inlet, outlet with combustion mouth, Can be pre-vacuum and can pass hydrogen, argon, nitrogen, oxygen, carbon monoxide, ammonia decomposition gas and other gases, The furnace has the advantages of temperature field equilibrium, low surface temperature, fast lifting temperature rate, energy saving and so on, It is an ideal product for universities, scientific research institutes and industrial and mining enterprises to do atmosphere protection, sintering and atmosphere reduction.
Software control system:
Equipped with communication interface and software, you can directly control the parameters of the furnace through the computer, and can observe the PV and SV temperature value of the furnace and the instrument operation from the computer. The actual heating curve of the stove will be drawn in real time, and can save the temperature data at each moment, which can be adjusted at any time.(pick out and buy)
1.heating element
(1)The thermal end thickness of silicon molybdenum rod D1 is 7mm, 6MM used by general manufacturers, the heating end is thick, not easy to bend, the surface load of the rod is low, and long service life.
(2)The use of hot bending technology, uniform resistance and long service life
(3)The length of Lu cold end is 210mm, the 400X300X300 furnace is 260, generally manufacturers use 180mm, the cold end long furnace roof temperature is low.
(4)The surface oxidized amorphous film of the silicon molybdenum rod is dense, which can better protect the silicon molybdenum rod.
2.Install the air switch and the leakage protector:
When the circuit is overcurrent or leakage, the empty meeting is automatically disconnected.
3.Can be connected to the computer
The furnace is equipped with communication interface and software, which can directly control the parameters of the furnace through the computer, and can observe the temperature value of PV and SV on the furnace and the instrument from the computer. The actual heating curve of the furnace will be drawn in real time, and can save the temperature data of each moment, which can be adjusted at any time
4.Furnace
(1)vacuum suction filter formed alumina polycrystalline fiber curing furnace.
(2)Using Japanese technology to shape.
(3)The spacing and pitch of the resistance wire in the furnace are all arranged according to the Japanese thermal technology and simulate the thermal field through the thermal software.
(4)With 4 weeks of heating, the temperature field is more balanced.
(5)The furnace does not drop powder, the material insulation performance is good, high reflectivity, temperature field balance, resistance to heat rise and cold contraction ability is strong.
5.Optional UL certification electrical board (all imported electrical appliances approved by UL certification)

KBF1700-HAF hydrogen atmosphere furnace technical parameters:

Modelpowerfurnace size mmshape dimension mmvoltage phasenumber of phases
KBF1700-HAF-14KW160*150*150550*470*810220Vsingle phase
KBF1700-HAF-28KW 300*200*200600*610*860220Vsingle phase
KBF1700-HAF-316KW400*300*300700*710*1355 380V single phase

Electrical appliances: Zhejiang Chint
Si: 106 / 16E Simenon, Germany
Maximum temperature: 1,600℃
Rated temperature: 1,550℃
Heating rate: 20℃ / Min (can be modified upon request)
Recommended heating rate: 10℃ / Min
Thermocouple model: Type B
6.Trigger
1.Phase shift trigger (with soft start and flow function)
2.Current limiting function: No matter how the heating rate is set, the secondary current of our electric furnace is always limited to 170A, which effectively protects the equipment. Due to the flow limiting function, the furnace door can be opened at high temperature. If the electric furnace you use needs to limit the heating rate of the furnace (e. g., specified as not exceeding 3℃ / min, etc., after 1400℃), a non-current limiting trigger is used. If the heating rate is set correctly or the heating rate is too fast, the current will be very large and the equipment will trip or burn out.
3.Soft start function: the current will slowly increase when the machine is started, and the impact on the heating element is small, which is conducive to the extension of the life of the heating element
4.protection function: if the current exceeds the limit current, the trigger will automatically cut off the trigger function to cut off the current protection electric furnace.
7.UL certified transformer
1.Transformer cable is pure copper
2.The transformer sound is small
3.the transformer temperature and body is low
8.Certification

Scroll to Top