OTL1200-RTP rapid heating furnace

Product name: OTL1200-RTP rapid heating furnace

Model number: OTL1200-RTP

Product Summary:
OTL1200-RTP rapid heating tube furnace is open, the furnace body adopts double shell structure, the furnace adopts Japanese imported alumina polycrystal fiber material, because the maximum heating rate can reach 50℃ / s, the substrate film can be quickly returned processing, the furnace is equipped with a travel switch, automatically cut off the power supply when the furnace cover is opened.

product description
OTL1200-RTP rapid heating furnace detailed introduction:

RTP1200 Fast heating tube furnace is open type, the furnace body adopts double shell structure, the furnace adopts Japan imported alumina polycrystal fiber material, because the maximum heating rate can reach 50℃ / s, can the substrate film can be quickly returned processing, the furnace is equipped with a travel switch, when the furnace cover is opened will automatically cut off the power supply.
1.Install the air switch and the leakage protector:
When the circuit is overcurrent or leakage, the empty meeting is automatically disconnected.
2.You can connect to the computer
The furnace is equipped with communication interface and software, which can directly control the parameters of the furnace through the computer, and can observe the temperature value of PV and SV on the furnace and the instrument from the computer. The actual heating curve of the furnace will be drawn in real time, and can save the temperature data of each moment, which can be adjusted at any time

3.Furnace
(1)vacuum suction filter molding of high quality high purity alumina polycrystalline fiber curing furnace.
(2)Using Japanese technology to shape.
(3)The spacing and pitch of the resistance wire in the furnace are all arranged according to the optimal thermal technology in Japan, and the thermal field is simulated through the thermal software
(4)Using 4 weeks of heating makes the temperature field more balanced
4.Optional UL certification electrical board (all imported electrical appliances approved by UL certification)
OTL1200-RTP rapid heating furnace technical parameters:
Model number: OTL1200-RTP
Power: 12KW
Furnace tube size (OD) (mm): Dia 100 (OD) X600mm
Power supply voltage: 380V
Phase number: single phase
Heating element: 12 iodine tungsten lamp tubes
Control accuracy: + / -1℃
Trigger: Phase shift trigger
Electrical appliances: Zhejiang Chint
SiriCR: 106 / 16E Simenon, Germany
Maximum temperature: 1,000℃
Heating section length: 240mm
Heating rate: 50℃ / S (modified upon request)
Thermocouple model: Model K type
Outer surface temperature: 45℃

Temperature control mode
Adopt temperature control instrument (standard)
1.Intelligent PID adjustment of temperature control in section 30.
2with over temperature protection when the electric furnace heating circuit automatically over temperature or break off, (when the electric furnace temperature exceeds 1020 degrees or the thermocouple burned off, the AC relay on the main circuit will automatically disconnected, the main circuit disconnected, the panel ON lamp off, OFF light, limited protection electric furnace).
3.with 485 communication interface (standard when choosing software)
4.with the function of power off protection, that is, when the power off to start the furnace, the program does not rise from the starting temperature, but from the furnace temperature when the power off.
5.The instrument has the function of temperature self-setting